5141 / 5142 / 5144 molbox RFM Gas Flow Calibrator
Kits
靖江中諾儀器有限公司為福祿克中國總代理,歡迎來電洽談!
High end gas flow calibration made simple
Fluke Calibration 5141, 5142 and 5144
molbox RFM gas flow calibrator kits feature molbox RFM and molbloc-L
configurations optimized to cover a very wide workload combined with molstic-L
and other interconnect hardware needed for a complete calibration system.
Simply supply 90 psi of pure N2 or air to the system. An included regulator
delivers a stable regulated flow to the molbloc. Downstream of the molbloc, an
included metering valve allows you to regulate the mass flow value required for
the DUT. It’s that simple; supply 90 psi gas upstream, tare the reading, and
adjust the fine metering valve to deliver the desired mass flow rate with an
accuracy of ± 0.5 % of reading. Fluke Calibration gas flow calibrator kits
eliminate complication, but don’t compromise molbox/molbloc best in class performance.
·Flow ranges cover from 1 sccm to 50 slm
·± 0.5 % of
reading total one-year accuracy
·Measures mass and volume flow with user settable reference pressure
and temperature conditions
·Reference standard is upstream of the device under test, eliminating
cross contamination and expensive damage to the flow elements
·Integrated gas flow regulation and adjustment hardware
·Expandable to over 5000 slm with additional molbloc flow elements
and control hardware
·Includes traceable calibration for operation in N2 and Air, with
corrections for other gases
·Gas supply and test adapters included for 1/4 in tube, 1/4 in NPT
and 1/4 in BSP
·Includes advanced functions such as totalize, average, hi/lo,
deviation, on-board purge, leak test and tare, available from the front control
panel or by RS-232 and IEEE-488 remote interfaces
A new standard in high end gas flow
calibration
Fluke Calibration revolutionized high end
gas flow calibration with the introduction of molbox/ molbloc mass flow
standards. molbox/molbloc systems have replaced large, inflexible, error prone
volumetric piston provers and bell provers with a compact, easy to use,
versatile digital standard. molbox/molbloc systems are the standard of choice
for high end calibration labs and mass flow device manufacturers around the
world. Improvements offered by molbox/molbloc include:
·True mass flow calibration and traceability— no
corrections from volumetric to mass
·Real-time digital mass flow indication, loaded with features—easily
automated
·No moving parts—uninterrupted gas flow measurement, no fluctuations from piston
stroke
·Flexibility to be positioned upstream or downstream— can
calibrate at many line pressures
·Modular components allows future upgrade and expansion
·Very wide range with small footprint—no flow
straightening hardware required
Whether your requirement is manual calibration
of a simple variable area flow meter (rotameter) or fully automated calibration
of a mass flow controller (MFC), molbox/molbloc systems offer the ideal
solution. A molbox RFM with one or more molbloc flow elements can cover a wide
range of flow calibration devices with total one-year measurement uncertainty
of ± 0.5 % of reading. If your needs change, molbloc flow elements can
be added, with models covering flows from 1 sccm to over 5000 slm (175 scfm).
If better measurement uncertainty is needed, the same flow elements can be used
with molbox1+ to achieve accuracy as good as ± 0.125 % of reading.
靖江中諾儀器有限公司為福祿克中國總代理,歡迎來電洽談!
Measurement specifications
|
Gas
calibration included
|
Nitrogen
(N2), Air
|
Maximum
working measurement line pressure
|
600 kPa
(87 psi) absolute
|
Measurement
range
|
<1sccm
to 50 slm, depending on model. Flow ranges depend on test gas. Ranged
indicated for N2 and Air.
|
General specifications
|
Power
requirements
|
85 V ac
to 264 V ac, 47 Hz to 440 Hz, 18 VA max consumption
|
Normal operating
temperature range
|
15 °C to
30 °C (59 °F to 86 °F)
|
Storage
temperature range
|
-20 °C
to 70 °C (-4 °F to 158 °F)
|
Vibration
|
Meets
MIL-T-28800D
|
molbox
RFM weight and dimwnsions
|
2.55 kg (5.6
lb) max
8 cm x 22.5 cm
x 20 cm (3.1 in x 8.9 in x 7.9 in) approx
|
molstic
with molbloc weight and dimensions (max)
|
9 kg (20 lb)
19 cm x 81.28
cm x 15.25 cm (7.5 in x 32 in x 6 in) approx
|
Supply
gas required
|
99.998% pure N2 or
Air regulated to 90 psig (600 kPa)
Note: a fine
pressure regulator is included, but an additional regulator (not included)
may be required to reduce pressure supplied to the 514X kit to approximately
90 psi
|
Gas
calculations reported
|
Nitrogen
(N2), Air, Argon (Ar), Carbon Monoxide (CO), Helium (He), Oxygen
(O2),
Carbon Dioxide (CO2), Carbon Tetrafluoride (CF4),
Ethane (C2<;/sub>H6), Ethylene
(C2H4), Fluoroform (CHF3), Hexafluoroethane (C2F6), Hydrogen (H2), Methane
(CH4),
Nitrous Oxide (N2O), Propane (C3H8), Sulfur Hexafluoride (SF6)
|
Flow
connections for gas supply and device under test
|
1/4 in
tube with adapters included for 1/4 in NOT and 1/4 in BSP
|
靖江中諾儀器有限公司為福祿克中國總代理,歡迎來電洽談!